Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Chapter IV
Calibration
Valve Reference Guide:
Figure IV-1
.represents .the .valve .positioning .for .the .Thermo .Scientific .
Orion .2111LL .sodium .monitor .during .measurement .and .calibration .
modes . . .This .quick .reference .tool .indicates .the .proper .position .of .the .4 .
main .valves .used .when .starting .calibrating .the .instrument . .
• . Calibration .shut .off .valve
• . Left .air .valve
• . Right .air .valve
• . Diverter .valve
Each .2111LL .is .shipped .with .a .Valve .Reference .Guide .as .a .self-adhesive .
tool .to .mount .either .to .the .back .PVC .panel .or .on .the .inside .of .the .2111LL .
enclosure .(Cat . .No . .2111EN) . . .
Using the Valve Reference Guide:
Measure Mode:
Turn valves to black arrow position.
Calibrate Mode:
Turn valves to red arrow position.
Flow Regulation:
Turn valves indicated in blue to either Open or Close
position for sample flow adjustment.
Positioning of Valves
for Calibration
Figure IV-