II-5
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
8 . . Invert .electrolyte .bottle .and .snap .into .clip . . .Refer .to .
Figure II-14
. . .Use .
pushpin .supplied .to .puncture .three .air .vents .on .bottom .of .bottle . .
Caution:
Failure .to .vent .fill .solution .bottle .will .lead .to .noisy .and .
drifting .output .signals . .
s
9 . . Plug .the .electrode .cable .marked .“Reference” .into .the .top .of .the .
electrode, .and .tighten .the .screw .cap . . .Be .sure .to .push .back .black .cap .
to .verify .a .secure .connection .between .male .and .female .pin .connection .
prior .to .tightening .
10 . .Tighten .screw .cap .connection .to .cable . .
Note:
Do .not .twist .cables .while .tightening .connection . . .Twisting .may .
cause .damage .requiring .premature .replacement .of .the .cables .
11 . .Insert .the .reference .electrode .into .its .port .in .the .flow .cell .cap . . .Refer .
Figure II-12
.for .location . .
Figure II-4
Reference Mounting Clip
Air Vents
Bottle
Clip (on Panel)