II-
Instrument Preparation
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Refer .to .
Appendix A1, Mounting Dimensions.
Warning:
Do .not .connect .power .prior .to .mounting .and .plumbing .of .the .
instrument . . .
s
.
• . Select .a .site .for .the .instrument .that .allows .it .to .be .permanently .bolted .
with .ample .height .for .atmospheric .drain .operation . . .Be .sure .there .is .
ready .access .to .both .electronic .controls .and .flow .cell .calibration .port . .
• . A .clearance .of .15 .inches .(about .40 .cm) .must .be .allowed .above .the .
calibration .port . . .Insert .the .pipette .vertically .(not .angled) .during .the .
calibration . .
• . The .analyzer .location .must .permit .connection .to .a .plumbed .in .sample .
line, .a .drain, .and .AC .power .supply .and .any .connections .for .output .
devices . .
• . The .analyzer .should .be .mounted .as .close .to .the .sampling .point .as .
possible, .this .ensures .the .fastest .possible .response .to .a .changing .sample .
condition . . .Refer .to .
Appendix A-4, Sample Conditions
.
• . For .proper .flow .cell .operations, .instrument .must .be .installed .straight .
and .level .upon .its .mounting .location . . .Failure .to .level .may .cause .poor .
siphoning .in .flow .cell .
1 . . Prepare .mounting .holes . . .Carefully .lift .the .analyzer .and .bolt .into .place . .
Do .not .lift .the .instrument .by .holding .on .to .any .of .the .plumbing .or .
fluid .handling .components .
2 . . Connect .a .waste .line .to .the .outlet .of .the .instrument, .which .is .3⁄4 .NPT .
male . . .The .waste .line .should .be .connected .to .a .drain .of .sufficient .
capacity, .0 .5 .inch .(1 .27 .cm) .OD .is .recommended .
3 . . Connect .a .sample .line .to .the .inlet .of .the .instrument, .which .is .1⁄4 .NPT .
female . . .It .is .recommended .that .a .shut-off .valve .be .installed .at .the .
sampling .point .
4 . . The .system .must .be .mounted .and .leveled, .vertically .for .proper .
operation .
Mounting and
Plumbing Instructions
Recommendations
Instructions