IV-
Calibration
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Stage
User Action
Marquee
Main Display
top line
bottom line
Valve position / Notes
Stability Timeout
Error 4
Decide to continue calibration
despite instability or abort
READINGS WERE
UNSTABLE. PRESS
ENTER TO CONTINUE
CALIBRATION OR
PRESS EXIT TO ABORT
E041
Stability was not reached before
timeout
If
is pressed system will
continue from step where stability
timed out.
If
is pressed refer to
Exit
or
Exit 4,
depending on the system
software revision
Calibration Error
Error 4
Review Marquee text to
understand the failure that
occurred
INVALID (NEGATIVE)
SLOPE
PRESS ENTER TO
ABORT CALIBRATION
E042
Invalid slope continue at
Exit 3
Error 09
Review Marquee text to
understand the failure that
occurred
BAD SLOPE
PRESS ENTER
E109
Bad Slope
Press
to continue at
stage 9
Error 5
Review Marquee text to
understand the failure that
occurred
BAD E
0
PRESS ENTER
E125
Bad E
0
Press
to continue at
stage 9