VI-
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
• . Displays .up .to .the .last .five .calibrations .using .three .information .screens:
1 . . Slope .and .E
0
2 . . Date .and .Time
3 . . Temperature .and .Electrode .Response .Time
• . Use .
.to .scroll .between .the .three .information .screens .for .one .
calibration
• . In .the .Date .and .Time .screen .use .
./ .
.to .view .up .to .the .last .five .
calibrations
Calibration Log
Menu
User Action
Marquee Scrolling
Main Display
top line
bottom line
Notes
Press
.
.
Calibration Log
CALIBRATION SLOPE AND E
0
##.# (Slope)
##.# (E
0
)
DIAGNOSTICS
appears as mode
Press
.
to display the calibration
time and date.
LAST CALIBRATION TIME/DATE
HH:MM
MM/DD or
DD/MM
In the calibration
time and date
display press
./ .
to scroll through
the last five
calibrations
Press
.
.
to display the calibration
temperature and electrode response.
CALIBRATION TEMP + ELECTRODE
RESPONSE TIME
##.# C (Temp)
### (Response)
Press
.
.
上海肯强贸易有限公司
http://www.kq17.com.cn