VI-5
Troubleshooting
Thermo Scientific Orion 2111LL Sodium Monitor User Guide
ROSS and the COIL tradedress are trademarks of Thermo Fisher Scientific Inc. US Patent 6,793,787
Displayed
Event Code
Description
Cause
Troubleshooting
System Event/Error Codes - Yellow LED
E004
DEFAULT VALUES
Instrument has been reset or is new
out of box
Calibrate the instrument
E0
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E0
TEMP OUT OF RANGE
Temperature sensor on channel 1 out
of range (5.1 ˚C to 45 ˚C)
Check ATC cable and probe connections. Replace if needed.
E
CAL DUE
User calibration required. No
calibration has been performed in
user specified time interval.
Recalibrate the instrument.
E7
SOUT mA OVER RANGE
Measured sodium value above Sout
high limit value
Verify customer selectable Sout high limit value in setup menu
E8
SOUT mA UNDER RANGE Measured sodium value below Sout
low limit value
Verify customer selectable Sout low limit value in setup menu
E9
TOUT mA OVER RANGE
Measured temperature value above
Tout high limit value
Verify customer selectable Tout high limit value in setup menu
E30
TOUT mA UNDER RANGE Measured temperature value below
Tout low limit value
Verify customer selectable Tout low limit value in setup menu
System Error Codes - Red LED
E001
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E002
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E005
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E006
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E018
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E019
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E020
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E022
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E024
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E026
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E027
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E029
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E030
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
E038
HARDWARE FAULT
System Error
Contact Customer Support for Orion products 1-800-225-1480
上海肯强贸易有限公司