5
Diagnostics
Monitoring the Pump Pulsation
44
Accela Pump Hardware Manual
Thermo Scientific
Using Xcalibur to Monitor the Pump Status
You can find information about the pressure variation (absolute pressure and standard
deviation of the pressure) on the Status page of the Information view. This view appears on
the left side of the Home Page window.
If you cannot see the Information view, it has been turned off.
Y
To open the Info view and check the status of the Accela Pump
1. From the Home Page window, choose
View > Info View
to turn the Information view on
and off.
2. To display the Status page, click the
Status
tab.
3. To display the status pages for the Accela Pump, click the
Accela Pump
listing in the
directory tree.
Check the value for the standard deviation of the pressure listed on the General page (see
Figure 32
).
Figure 32.
Status pages for the Accela Pump
Pressure status readback