A
Discontinued Versions of the Accela Pump
Installing the Check Valves in the Discontinued LDA
Thermo Scientific
Accela Pump Hardware Manual
103
Figure 74.
Installing the PEEK film seals and discontinued check valves
2. Insert the lower displacement chamber into the LDA body just far enough so that the
body serves as a vice, and then tighten the check valve fitting with an 8 mm open-ended
wrench (see
Figure 75
).
3. Remove the displacement chamber from the LDA body. Rotate it and repeat
step 2
.
Figure 75.
Tightening the outlet check valve fitting
Inlet
check
valve
Outlet
check
valve
PEEK film seals
Lower displacement chamber