Preparing the Pump for Operation
Using the Data System to Start and Stop the Solvent Flow
40
Accela Pump Hardware Manual
Thermo Scientific
9. Click
(Stop) to stop the solvent flow.
10. If you opened the purge valve to let the solvent flow bypass the LC column, close the
purge valve by turning the wingnut clockwise.
Using ChromQuest to Start and Stop the Solvent Flow
You can start and stop the solvent flow by using the Direct Control area of the Instrument
Status page for the Accela Pump.
Y
To start and stop the solvent flow by using the ChromQuest data system
1. From the Main Menu window, double-click the icon for your instrument.
The online Instrument window opens.
2. Choose
Control > Instrument Status
.
The Instrument Status window appears with the Accela Pump page displayed (see
Figure 31
). Use the Accela Pump page to view the pump status or to start an isocratic
solvent flow.
Figure 31.
Accela Pump page of the Instrument Status window
3. In the Flow box, type
1000
.
4. Type percentages for the inlet (solvent) lines through which you want to draw solvent.
5. Click
Set%
.