Making the Initial Back Panel Connections
Thermo Scientific
Accela Pump Hardware Manual
15
Connecting the System Interconnect Cable
The system interconnect cable synchronizes the timing of the Accela devices. The
interconnect cable has seven combicon connectors: three are labeled DETECTOR; two are
labeled PUMP; one has a small, A/S tag on its adjacent cable; and one has a small, M/S tag on
its adjacent cable (see
Figure 8
).
Figure 9
shows the connections to the system modules.
Figure 8.
System interconnect cable with seven combicon connectors
Y
To connect the system interconnect cable
1. Plug one of the pump connectors with a yellow PUMP sticker into the 8-pin socket on
the back panel of the Accela Pump.
2. Plug the autosampler connector with the A/S tag on its adjacent cable into the left, 8-pin
socket on the back panel of the Accela Autosampler.
3. If your LC system contains an Accela PDA Detector, plug one of the detector connectors
with a blue DETECTOR sticker into the left, 8-pin socket on the back panel of the
detector.
IMPORTANT
You can use the older 5-connector version of the system interconnect cable
with Accela LC system. If you use the cable that has 5-combicon connectors, attach the
connector that has a blue, LC PUMP label to the Accela Pump.