Discontinued Versions of the Accela Pump
Installing the Components of the Discontinued LDA Stack
104
Accela Pump Hardware Manual
Thermo Scientific
Installing the Components of the Discontinued LDA Stack
If they are installed correctly, the inner components of the LDA form a perfect stack that
ensures smooth movement of the pistons.
Y
To install the inner components of the LDA stack
1. Because the fit between the check valve holders an the lower displacement chamber is
loose, place the LDA on a flat surface, and then insert the lower displacement chamber
and check valve assembly. The orientation arrow on the lower displacement chamber
points away from the LDA body (see
Figure 76
).
2. Install the upper displacement chamber and two PEEK film seals.
3. Install the outlet module.
Figure 76.
Installing the stacked components of the LDA
4. Slide the compression lid into the upper bracket of the LDA body (see
Figure 77
).
Figure 77.
Sliding the compression lid into the LDA body
Lower displacement
chamber
Upper displacement
chamber
Outlet module
PEEK sealing rings
LDA body
Compression lid