70 User Manual
Window Trigger
Window trigger provides a high trigger level and a low trigger level. The instrument
triggers when the input signal passes through the high trigger level or the low trigger level.
There are two kinds of window types: Absolute and Relative. They have different trigger
level adjustment methods. Under Absolute window type, the lower and the upper trigger
levels can be adjusted respectively via the Level knob; under Relative window type, adjust
the Center value to set the window center; adjust the Delta value to set the window range,
the lower and the upper trigger levels always move together.
Horizontal Trigger Position
High Level
Low Level
●
If the lower and the upper trigger levels are both within the waveform amplitude range,
the oscilloscope will trigger on both rising and falling edge.
●
If the upper trigger level is within the waveform amplitude range while the lower
trigger level is out of the waveform amplitude range, the oscilloscope will trigger on
rising edge only.
●
If the lower trigger level is within the waveform amplitude range while the upper
trigger level is out of the waveform amplitude range, the oscilloscope will trigger on
falling edge only.
To set window trigger via Absolute window type:
1. Press the
Setup
button on the front panel to enter the TRIGGER function menu.
2. Press the
Type
softkey; then use the
Universal Knob
to select
Window
and push
down the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as the
trigger source.
4. Press the Window
Type
softkey to select Absolute.
5. Press the
Lower Upper
softkey to select
Lower
or
Upper
trigger level; then turn the
Trigger Level Knob
to adjust the position. The trigger level values are displayed at
the upper right corner of the screen.
The Lower trigger level cannot be higher than the upper trigger level. In the trigger
state message box,
L1
means the upper trigger level while
L2
means the lower trigger
level.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...
Page 205: ...User Manual 187 Figure 100 Option Information...