158 User Manual
To Change the Logic Threshold for Digital Channels
1. Press the
Digital
button on the front panel to open the DIGITAL function menu.
2. Press the
Threshold
softkey to enter the THRESHOLDS menu.
3. Press the
D0~D7
or
D8~D15
softkey, then turn the
Universal Knob
to select a
logic family preset or select Custom to define your own threshold.
Figure 83 THRESHOLDS Function Menu
Logic Family
Threshold Voltage
TTL
1.5V
CMOS
1.65V
LVCMOS3.3
1.65V
LVCMOS2.5
1.25V
Custom
Variable from -3V to +3V
●
The threshold you set applies to all channels.
●
Values greater than the set threshold are high (1) and values less than the set
threshold are low (0).
●
If the
D0~D7
or
D8~D15
softkey is set to
Custom
, press the
Custom
softkey, then
turn the
Universal Knob
to select the desired value and push the knob to
confirm.
To Reposition a Digital Channel
1. Press the
Digital
button on the front panel to open the DIGITAL function menu and
display the digital channel waveforms.
2. Press the
knob control softkey
then turn the
Universal Knob
to select the
channel that you wish to reposition. The selected waveform is highlighted in red.
3. Turn the
Universal Knob
to move the selected channel position.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...