30 User Manual
To Specify Amplitude Unit
Select the amplitude display unit for the current channel. The available units are
V
and
A
.
When the unit is changed, the unit displayed in the channel label will change accordingly.
1. Press
CH1
button on the front panel to enter the CH1 function menu.
2. Press the
Next Page
softkey to enter the second page of the CH1 function menu.
3. Press the
Unit
softkey to select the desired unit
V
or
A
.
The default setup is
V.
To Specify Deskew
The valid range of each analog channel is ±100ns.
1. Press
CH1
button on the front panel to enter the CH1 function menu.
2. Press the
Next Page
softkey to enter the second page of the CH1 function menu.
3. Press the
Deskew
softkey. Then turn the
Universal Knob
to change deskew.
To Invert a Waveform
When
Invert
is set to
On
, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed but keeps the original trigger settings.
Inverting a channel also changes the result of any math function selected and measure
function.
1. Press
CH1
button on the front panel to enter the CH1 function menu.
2. Press the
Next Page
softkey to enter the second page of the CH1 function menu.
3. Press the
Invert
softkey to turn on or off the invert display.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...
Page 205: ...User Manual 187 Figure 100 Option Information...