User Manual
169
Output
Connect AWG with USB cable, and the oscilloscope interface indicates "AWG device
connection success", Press the
Utility
→Next Page4/4→AWG, to enter the function menu
interface of the AWG system as shown below, including: Output, Wave, Setting, System,
Remove Device.
Figure 89 AWG Menu Interface
Press Output softkey to open or close the AWG output. When the output is on, the upper
right corner of the oscilloscope interface will display the corresponding waveform
frequency and amplitude information of AWG output wave.
Setting a Standard Waveform
1. To set standard waveform, sine wave example.
Press the
Utility
→Next Page4/4→AWG→Wave, Rotate the Universal knob to the Sine
wave. Then press the knob to select the interface shown as below:
Figure 90 Wave Type Interface
2. The following table shows the waveform type and its corresponding characteristics.
Set the waveform parameters for the required waveform by using the universal
knob and the menu system.