The display of waveform is ladder-like:
1)
The horizontal time base might be too low. Increase the horizontal time base to
increase the horizontal resolution and improve the display.
2)
If the display
Type
is “Vectors”, the lines between the sample points may cause
ladder-like display. Set
Type
to “Dots” to solve the problem.
7.
Failure to connect to the PC through USB:
Check the
IO Setting
in
Utility
to make sure that the setting in
USB Device
matches
the device currently connected. If changes are made then restart the oscilloscope for
the new settings to take effect (cycle the power).
8.
The USB storage device isn’t recognized:
1)
Check whether the USB storage device is working normally and formatted as
FAT32.
2)
Make sure that the USB interface is work normally. Test is with another USB
storage device.
3)
Make sure that the USB storage device being used is flash storage. This
oscilloscope does not support other hardware storage types.
4)
Restart the instrument and then insert the USB storage device to check it.
5)
If the USB storage device still cannot be used normally, please contact
Teledyne Test Tools
.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...
Page 205: ...User Manual 187 Figure 100 Option Information...