110 User Manual
LIN Triggering
This part will provide a brief introduction and description for the operation of the LIN
trigger.
The possible trigger conditions
●
Break
— The oscilloscope will be triggered at the position of break field’s break
delimiter.
●
ID
(Frame ID) — the oscilloscope will be triggered at the position of the identifier
field’s stop bit, if the value of a frame’s ID is equal to specified value.
(Note: If the data’s value is 0xXX, any data value will be matched)
●
ID + Data
(Frame ID and Data) — the oscilloscope triggers when a frame with an
ID and data equal to the selected values is detected. Use the Universal Knob to
select the value for the ID, Data1 and Data2.
1)
The ID’s value is the same as the value currently set.
2) If you have set either Data1’s or Data2’s value, and the signal has data that
matches that value. If you have set both Data1’s and Data2’s value, the signal
should has two consecutive data, the first data’s value is Data1, the second data
value is Data2.
(
Note
: If the data’s value is 0xXX, any data value will be matched)
●
Data Error
—the oscilloscope will be triggered when errors (such as ID check
error, checksum error, sync byte field error) are detected.
Operation steps:
1. Press
Setup
to enter the
TRIGGER
function menu.
2. Press
Type
and select
Serial
.
3. Press
Protocol
and select
LIN
.
4. Press
Trigger Setting
softkey to enter
LIN TRIG SET
menu.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...
Page 205: ...User Manual 187 Figure 100 Option Information...