50 User Manual
Use Sequence Mode
Sequence is also an acquisition mode, which does not display a waveform during the
sampling process. This improves the waveform capture rate, to a maximal capture rate of
400,000 wfs/s. So it can capture the small probability events effectively. The oscilloscope
runs and fills a memory segment for each trigger event. The oscilloscope screen is not
updated whilst the oscilloscope is busy acquiring multiple segments. The oscilloscope
continues to trigger until memory is filled, and then displays the waveforms on the screen.
To use the sequence mode, the HORIZONTAL Format must be set to
YT
.
Do the following steps to use the sequence mode.
1. Press the
Acquire
button on the front panel to enter the ACQUIRE function menu.
2. Press the
Sequence
softkey to enter the SEQUENCE function menu.
Figure 23 SEQUENCE Function Menu
3. Press the
Segments Set
softkey; and then turn the
Universal Knob
to select the
desired value.
Do the following steps to replay the sequence waveform under history mode:
1. Press the
History
softkey to enable HISTORY function.
Figure 24
HISTORY Function Menu
2. Press the
List
softkey to turn on the list display. The list records the acquisition time of
every frame and shows the frame number that is displaying on the screen.
3. Press the
Frame No.
softkey; and then turn the
Universal Knob
to select the frame to
display.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...
Page 205: ...User Manual 187 Figure 100 Option Information...