Note:
●
Signals with DC components or deviation would cause error or deviation of the FFT
waveform components. To reduce the DC components, set the Channel
Coupling
to
AC.
●
To reduce the random noise and aliasing frequency components of repetitive or
single pulse, set the
Acquisition
of the oscilloscope to
Average
.
10. Press the
Display
softkey to select
Split, Full Screen or Exclusive
display mode.
●
Split:
the source channel and the FFT operation results are displayed separately.
The time domain and frequency domain signals are displayed clearly.
●
Full Screen:
the source channel and the FFT operation results are displayed in
the same window to view the frequency spectrum more clearly and to perform
more precise measurements.
●
Exclusive:
the waveforms of channels are disabled, only the FFT operation
results are displayed in the window to view the frequency spectrum more clearly
and to perform more precise measurements.
Figure 59 FFT Waveform in Split Mode
11. Press the
Max Points
softkey to select the maximum number of points which are used
to compute the FFT waveform.
12. Press the
Autoset
softkey to automatically set the appropriate parameters for the FFT
measurement.
13. Press the
Max-Hold
softkey to turn on or turn off the Max Hold function.
Summary of Contents for T3DSO1000
Page 1: ...User Manual T3DSO1000 T3DSO1000A Series Digital Oscilloscope...
Page 62: ...44 User Manual Figure 16 x Interpolation Figure 17 Sinx x Interpolation...
Page 90: ...72 User Manual Figure 32 Relative Window Trigger...
Page 92: ...74 User Manual Figure 33 Interval Trigger...
Page 98: ...80 User Manual Universal Knob to select the desired value Figure 37 Pattern Trigger...
Page 118: ...100 User Manual Figure 50 UART Trigger...
Page 145: ...User Manual 127 function output Figure 63 Square Root...
Page 205: ...User Manual 187 Figure 100 Option Information...