Getting Started
Teledyne API – Model T300/T300M CO Analyzer
84
Z
ERO
/S
PAN
C
ALIBRATION
To perform the zero/span calibration procedure, press:
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the CO
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST > CAL SETUP
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST>
ZERO
CONC
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the CO
measurements.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=0.0 PPm CO= XX.XX
< TST
TST >
CAL SETUP
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the CO
measurement.
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls below
0.5 ppm.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST> ZERO
SPAN
CONC
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls below
0.5 PPM.
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XX.XX
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPM
CO=XX.XX
<TST TST> ENTR
CONC
EXIT
The
SPAN
button now
appears during the transition
from zero to span.
You may see both keys.
If either the
ZERO
or
SPAN
buttons fail to appear see the
Troubleshooting section for
tips.
Figure 3-29:
Zero/Span Calibration Procedure
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...