Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
107
5.4.4.
RANGE UNITS
The T300/T300M can display concentrations in parts per million (10
6
mols per mol,
PPM) or milligrams per cubic meter (mg/m
3
, MG). Changing units affects all of the
display, COMM port and DAS values for all reporting ranges regardless of the
analyzer’s range mode. To change the concentration units:
The following equations give
approximate conversions
between volume/volume units
and weight/volume units:
CO
CO ppb x 1.25 = CO ug/m
3
CO ppm x 1.25 = CO mg/m
3
SAMPLE
RANGE=50.0 PPM CO=XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
CONC UNITS:PPB
PPB PPM UGM MGM
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
RANGE MODE MENU
MODE
SET
UNIT
DIL
EXIT
SETUP X.X
RANGE MODE:SNGL
SNGL
DUAL AUTO
ENTR EXIT
Toggle these buttons
to select the units of
measure for the
reporting ranges.
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
In order to avoid a reference temperature bias, the analyzer must be
recalibrated after every change in reporting units.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Concentrations displayed in mg/m3 and ug/m3 use 0
° C
and 760 mmHg for
Standard Temperature and Pressure (STP). Consult your local regulations
for the STP used by your agency.
(Example: US EPA uses 25
o
C as the reference temperature).
Once the Units of Measurement have been changed from volumetric (ppb or
ppm) to mass units (µg/m3 or mg/m3) the analyzer MUST be recalibrated, as
the “expected span values” previously in effect will no longer be valid.
Simply entering new expected span values without running the entire
calibration routine IS NOT sufficient.
This will also counteract any discrepancies between STP definitions.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...