APPENDIX A-2: Setup Variables For Serial I/O
A-9
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
FILT_ASIZE Samples
48,
20
3, 8
,
40
20, 22
1–1000
Moving average filter size in
adaptive mode.
FILT_DELTA PPM
4,
0.7
5
15
3, 8
0.15
9, 12
0.4
19, 23
0.2
20, 22
1–1000
Absolute change to trigger
adaptive filter.
FILT_PCT %
10
5
20, 22, 23
1–100
Percent change to trigger
adaptive filter.
FILT_DELAY Seconds
90,
72
20, 22
0–180
Delay before leaving adaptive
filter mode.
FILT_ADAPT —
ON
ON,
OFF
ON
enables adaptive filter;
OFF
disables it.
CO2_DWELL
10
Seconds
0.1
0.1–30
Dwell time before taking each
sample.
CO2_FILT_ADAPT
10
— ON ON,
OFF
ON enables CO
2
adaptive filter;
OFF disables it.
CO2_FILT_SIZE
10
Samples
48
1–300
CO
2
moving average filter size.
CO2_FILT_ASIZE
10
Samples
12 1–300
CO
2
moving average filter size in
adaptive mode.
CO2_FILT_DELTA
10
% 2 0.01–10
Absolute CO
2
conc. change to
trigger adaptive filter.
CO2_FILT_PCT
10
%
10
0.1–100
Percent CO
2
conc. change to
trigger adaptive filter.
CO2_FILT_DELAY
10
Seconds
90 0–300
Delay before leaving CO
2
adaptive filter mode.
CO2_DIL_FACTOR
10
— 1
0.1–1000
Dilution factor for CO
2
. Used only
if is dilution enabled with
FACTORY_OPT
variable.
O2_DWELL
14
Seconds
1
0.1–30
Dwell time before taking each
sample.
O2_FILT_ADAPT
14
— ON
ON,
OFF
ON enables O
2
adaptive filter;
OFF disables it.
O2_FILT_SIZE
14
Samples
60
1–500
O
2
moving average filter size in
normal mode.
O2_FILT_ASIZE
14
Samples
10
1–500
O
2
moving average filter size in
adaptive mode.
O2_FILT_DELTA
14
% 2 0.1–100
Absolute change in O
2
concentration to shorten filter.
O2_FILT_PCT
14
%
2
0.1–100
Relative change in O
2
concentration to shorten filter.
O2_FILT_DELAY
14
Seconds
20 0–300
Delay before leaving O
2
adaptive
filter mode.
O2_DIL_FACTOR
14
— 1 0.1–1000
Dilution factor for O
2
. Used only if
is dilution enabled with
FACTORY_OPT
variable.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...