Teledyne API – Model T300/T300M CO Analyzer
Communications Setup and Operation
157
instruments are regarded as slaves of the host computer. The remote instruments are
unaware that they are connected to a multidrop bus and never initiate Hessen protocol
messages. They only respond to commands from the host computer and only when they
receive a command containing their own unique ID number.
The Hessen protocol is designed to accomplish two things: to obtain the status of remote
instruments, including the concentrations of all the gases measured; and to place remote
instruments into zero or span calibration or measure mode. API’s implementation
supports both of these principal features.
The Hessen protocol is not well defined, therefore while API’s application is completely
compatible with the protocol itself, it may be different from implementations by other
companies.
Note
The following sections describe the basics for setting up your instrument
to operate over a Hessen Protocol network.
For more detailed information as well as a list of host computer
commands and examples of command and response message syntax,
download the Manual Addendum for Hessen Protocol from the Teledyne
API web site: http://www.teledyne-api.com/manuals/.
6.7.2.1.
HESSEN COMM PORT CONFIGURATION
Hessen protocol requires the communication parameters of the T300/T300M Analyzer’s
COMM ports to be set differently than the standard configuration as shown in Table 6-4.
Table 6-4:
RS-232 Communication Parameters for Hessen Protocol
PARAMETER
STANDARD
HESSEN
Baud Rate
300 – 19200
1200
Data Bits
8
7
Stop Bits
1
2
Parity
None
Even
Duplex
Full
Half
To change the baud rate of the T300/T300M’s COMM ports, see Section 6.2.2.
To change the rest of the COMM port parameters listed in Table 6-4, see Section 6.2 and
Table 6-1.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Ensure that the communication parameters of the host computer are also
properly set.
Note
The instrument software has a 200 ms latency before it responds to
commands issued by the host computer. This latency should present no
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...