Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
227
9.7.2.
CO
2
SENSOR CALIBRATION PROCEDURE
Presented here are first the setup and then the calibration steps for the CO
2
Sensor.
9.7.2.1.
CO2 PNEUMATICS CONNECTIONS
The pneumatic connections for calibrating are as follows
Calibrated N
2
at
HIGH
Span
Concentration
Calibrated CO
2
at
16%
Span
Concentration
Source of
SAMPLE GAS
Removed during
calibration
SAMPLE
EXHAUST
PUMP
VENT
3-way
Valve
Manual
Control Valve
VENT
here if input
is pressurized
Instrument
Chassis
Figure 9-8:
CO
2
Sensor Calibration Set Up
CO
2
SENSOR ZERO GAS:
Teledyne API recommends using pure N
2
when calibration
the zero point of your CO
2
sensor option.
CO
2
SENSOR SPAN GAS:
Teledyne API recommends using 16% CO
2
in N
2
when
calibration the span point of your CO
2
sensor option (Table 3-12) is 20%.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...