Communications Setup and Operation
Teledyne API – Model T300/T300M CO Analyzer
146
6.2.2.
COM PORT BAUD RATE
To select the baud rate of either COM Port, go to SETUP>MORE>COMM and select
either COM1 or COM2 as follows (use COM2 to view/match your personal computer
baud rate when using the USB port, Section 6.6):
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
SETUP X.X COM1 MODE:0
<SET
SET>
EDIT
EXIT
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
COM1 BAUD RATE:115200
<SET
SET>
EDIT
EXIT
SETUP X.X
COM1 BAUD RATE:115200
PREV NEXT
ENTR
EXIT
Toggle these buttons
to cycle through the
available Baud rates:
300
1200
4800
9600
19200
38400
57600
115200
SETUP X.X
COM1 BAUD RATE:19200
PREV NEXT
ENTR
EXIT
Figure 6-2:
COMM Port Baud Rate
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...