Teledyne API – Model T300/T300M CO Analyzer
Troubleshooting and Service
263
12.1.4.
STATUS LEDS
Several color-coded light-emitting diodes (LEDs) are located inside the instrument to
assist in determining if the analyzer’s CPU, I
2
C bus and relay board, GFC Wheel and the
sync/demodulator board are functioning properly.
12.1.4.1.
MOTHERBOARD STATUS INDICATOR (WATCHDOG)
DS5, a red LED, that is located on upper portion of the motherboard, just to the right of
the CPU board, flashes when the CPU is running the main program loop. After power-
up, approximately 30 to 60 seconds, DS5 should flash on and off. If characters are
written to the front panel display but DS5 does not flash then the program files have
become corrupted. If after 30 – 60 seconds neither the DS5 is flashing or no characters
have been written to the front panel display then the CPU is bad and must be replaced.
Motherboard
CPU Status LED
Figure 12-3:
CPU Status Indicator
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...