Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
APPENDIX A-3: Warnings and Test Functions
A-23
Table A-3: T300/T300M and M300E/EM Test Functions, Revision L.8
TEST FUNCTION NAME
MESSAGE TEXT
DESCRIPTION
RANGE RANGE=50.0
PPM
3
CO RANGE=50.0 PPM
3, 7
D/A range in single or auto-range modes.
RANGE1 RANGE1=50.0
PPM
3
CO RANGE1=50.0 PPM
3, 7
D/A #1 range in dual range mode.
RANGE2 RANGE2=50.0
PPM
3
CO RANGE2=50.0 PPM
3, 7
D/A #2 range in dual range mode.
CO2RANGE
CO2 RANGE=20 %
7
CO
2
range.
O2RANGE
O2 RANGE=100 %
10
O
2
range.
STABILITY STABIL=0.0
PPM
3
CO STB=0.0 PPM
3, 7, 10
CO2 STB=0.0 %
7
O2 STB=0.0 %
10
Concentration stability (standard deviation
based on setting of
STABIL_FREQ
and
STABIL_SAMPLES
).
RESPONSE
2
RSP=0.20(0.00)
SEC Instrument response. Length of each
signal processing loop. Time in
parenthesis is standard deviation.
COMEAS
CO MEAS=4125.0 MV
Detector measure reading.
COREF
CO REF=3750.0 MV
Detector reference reading.
MRRATIO
MR RATIO=1.100
Measure/reference ratio.
AUTOZERO
4, 5
AZERO
RATIO=1.234 Measure/reference ratio during auto-
reference.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
PURGEPRESS
9
PURGE=7.5 PSIG
Purge pressure
VACUUM
8
VAC=6.8
IN-HG-A Vacuum pressure.
SAMPFLOW
6
SAMP FL=751 CC/M
Sample flow rate.
SAMPTEMP
SAMPLE TEMP=26.8 C
Sample temperature.
BENCHTEMP BENCH
TEMP=48.1
C Bench
temperature.
WHEELTEMP
WHEEL TEMP=68.1 C
Wheel temperature.
O2CELLTEMP
10
O2 CELL TEMP=50.2 C
O
2
sensor cell temperature.
BOXTEMP
BOX TEMP=26.8 C
Internal box temperature.
BOXTEMP2
4
BOX TEMP2=29.6 C
Internal box temperature #2.
OVENTEMP
11
OVEN TEMP=30.1 C
Oven temperature
PHOTOTEMP
PHT DRIVE=2500.0 MV
Photometer temperature.
COSLOPE SLOPE=1.000
CO SLOPE=1.000
7
CO slope for current range, computed
during zero/span calibration.
COSLOPE1 SLOPE1=1.000
CO SLOPE1=1.000
7
CO slope for range #1 in dual range
mode, computed during zero/span
calibration.
COSLOPE2 SLOPE2=1.000
CO SLOPE2=1.000
7
CO slope for range #2 in dual range
mode, computed during zero/span
calibration.
COOFFSET OFFSET=0.000
CO OFFSET=0.000
7
CO offset for current range, computed
during zero/span calibration.
COOFFSET1 OFFSET1=0.000
CO OFFSET1=0.000
7
CO offset for range #1 in dual range
mode, computed during zero/span
calibration.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...