Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
219
9.6.
CALIBRATION OF THE T300/T300M’S ELECTRONIC
SUBSYSTEMS
9.6.1.
DARK CALIBRATION TEST
The dark calibration test interrupts the signal path between the IR photo-detector and the
remainder of the sync/demod board circuitry. This allows the instrument to compensate
for any voltage levels inherent in the sync/demod circuitry that might effect the
calculation of CO concentration.
Performing this calibration returns two offset voltages, one for
CO MEAS
and one for
CO REF
that are automatically added to the CPU’s calculation routine. The two offset
voltages from the last calibration procedure may be reviewed by the user via the front
panel display. (See also Section 5.9.5).
To activate the dark calibration procedure or review the results of a previous calibration,
press:
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...