Theory of Operation
Teledyne API – Model T300/T300M CO Analyzer
314
The four sample and hold circuits are designated as follows:
Table 13-2:
Sync DEMOD Sample and Hold Circuits
Designation
Active When:
IR BEAM PASSING THROUGH
Segment Sensor Pulse is:
Measure Gate
MEASUREMENT cell of GFC Wheel
HIGH
Measure Dark Gate
MEASUREMENT Cell of GFC Wheel
LOW
Reference Gate
REFERENCE cell of GFC Wheel
HIGH
Reference Dark Gate
REFERENCE cell of GFC Wheel
LOW
Timing for activating the Sample and Hold Circuits is provided by a Phase Lock Loop
(PLL) circuit. Using the segment sensor output as a reference signal the PLL generates
clock signal at ten times that frequency. This faster clock signal is used by the PLD to
make the Sample and Hold Circuits capture the signal during the center portions of the
detected waveform, ignore the rising and falling edges of the detector signal.
Sample & Hold
Active
Detector
Output
Sample & Hold
Inactive
Figure 13-14: Sample & Hold Timing
13.4.3.2.
SYNC/DEMOD STATUS LEDS
The following two status LEDs located on the sync/demod board provide additional
diagnostic tools for checking the GFC Wheel rotation.
Table 13-3:
Sync/Demod Status LED Activity
LED
Function
Status OK
Fault Status
D1
M/R Sensor Status
LED flashes approximately
2/second
LED is stuck
ON or OFF
D2
Segment Sensor Status
LED flashes approximately
6/second
LED is stuck
ON or OFF
See Section 12.1.4.2 for more information.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...