APPENDIX A-2: Setup Variables For Serial I/O
Teledyne API - T300/T300M and M300E/EM PN 04906H (DCN5840)
A-18
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
SERIAL_NUMBER —
“00000000 ”
0
Any character
in the allowed
character set.
Up to 100
characters
long.
Unique serial number for
instrument.
DISP_INTENSITY —
HIGH
0
HIGH,
MED,
LOW,
DIM
Front panel display intensity.
I2C_RESET_ENABLE — ON OFF,
ON
ON
enables automatic reset of
the I
2
C bus in the event of
communication failures;
OFF
disables automatic reset.
CLOCK_FORMAT —
“TIME=%H:%
M:%S”
Any character
in the allowed
character set.
Up to 100
characters
long.
Time-of-day clock format flags.
Enclose value in double quotes
(") when setting from the RS-232
interface.
“%a” = Abbreviated weekday
name.
“%b” = Abbreviated month name.
“%d” = Day of month as decimal
number (01 – 31).
“%H” = Hour in 24-hour format
(00 – 23).
“%I” = Hour in 12-hour format (01
– 12).
“%j” = Day of year as decimal
number (001 – 366).
“%m” = Month as decimal
number (01 – 12).
“%M” = Minute as decimal
number (00 – 59).
“%p” = A.M./P.M. indicator for
12-hour clock.
“%S” = Second as decimal
number (00 – 59).
“%w” = Weekday as decimal
number (0 – 6; Sunday is 0).
“%y” = Year without century, as
decimal number (00 – 99).
“%Y” = Year with century, as
decimal number.
“%%” = Percent sign.
ALARM_TRIGGER
3,4
Cycles
10 1–100
Concentration alarm trigger
sensitivity adjustment.
REF_SDEV_LIMIT mV
50
0.1–500
Reference detector standard
deviation must be below this limit
to switch out of startup mode.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...