APPENDIX A-2: Setup Variables For Serial I/O
A-17
Setup Variable
Numeric
Units
Default
Value
Value
Range
Description
BENCH_DERIV —
2
0
19,23
0–100
100V optical bench temperature
PID derivative coefficient.
BENCH_PROP2 1/ºC
5
0.75
19,23
0–100
200V optical bench temperature
PID proportional coefficient.
Proportional band is the
reciprocal of this setting.
BENCH_INTEG2 —
0.5
0.75
19,23
0–100
200V optical bench temperature
PID integral coefficient.
BENCH_DERIV2 —
2
0
19,23
0–100
200V optical bench temperature
PID derivative coefficient.
WHEEL_CYCLE Seconds
4
2
4,9,12,18
8
19,23
0.5–30
Wheel temperature control cycle
period.
WHEEL_PROP 1/ºC
1
0.3
19,23
0–100
100V wheel temperature PID
proportional coefficient.
Proportional band is the
reciprocal of this setting.
WHEEL_INTEG —
0.135
0.035
4,9,12,18
0.06
19,23
0–100
100V wheel temperature PID
integral coefficient.
WHEEL_DERIV —
2
0
19,23
0–100
100V wheel temperature PID
derivative coefficient.
WHEEL_PROP2 1/ºC
1
0.1
19,23
0–100
200V wheel temperature PID
proportional coefficient.
Proportional band is the
reciprocal of this setting.
WHEEL_INTEG2 —
0.135
0.035
4,9,12,18
0.01
19,23
0–100
200V wheel temperature PID
integral coefficient.
WHEEL_DERIV2 —
2
0
19,23
0–100
200V wheel temperature PID
derivative coefficient.
O2_CELL_CYCLE
14
Seconds
10 0.5–30
O
2
cell temperature control cycle
period.
O2_CELL_PROP
14
— 1 0–10
O
2
cell PID temperature control
proportional coefficient.
O2_CELL_INTEG
14
— 0.1 0–10
O
2
cell PID temperature control
integral coefficient.
O2_CELL_DERIV
14
— 0
(disabled)
0–10
O
2
cell PID temperature control
derivative coefficient.
BOX_TEMP_GAIN PPB/DegC
0,
5
9
0–100
Gain factor for box temperature
compensation of concentration.
TPC_ENABLE —
ON
OFF,
ON
ON
enables temperature/
pressure compensation;
OFF
disables it.
CONC_LIN_ENABLE — ON OFF,
ON
ON
enables concentration
linearization;
OFF
disables it.
STAT_REP_PERIOD
17
Seconds
1
0.5–120
TAI protocol status message
report period.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...