Teledyne API – Model T300/T300M CO Analyzer
EPA Calibration Protocol
239
For each mid point:
SPAN CAL M
A1:CONC1=50 PPM CO = XXXX
< TST TST >
ZERO
SPAN
CONC
EXIT
Wait until
STABIL
falls
below
0.2 PPM
(for M300E).
This may take
several minutes.
SAMPLE
COSTB=XXXX PPB
CO=XXXX
< TST TST > CAL
CALZ
CALS
SETUP
Record the CO
reading as
displayed on the
instrument’s front
panel.
Press
EXIT
to
Return to the
Main
SAMPLE
Display.
ACTION:
Allow Calibration Gas diluted to
proper concentration for
Midpoint N+1 to enter the sample
port.
ACTION:
Allow calibration gas diluted to proper concentration for
Midpoint N to enter the sample port
SAMPLE A1:CONC1=50 PPM CO = XXXX
< TST
TST
>
CAL
SETUP
SAMPLE
CO STB=XXXX PPB
CO=XXXX
< TST TST >
CAL
SETUP
Set the Display to show the
COSTB
test function.
This function calculates the
stability of the CO
measurement.
Plot the analyzer responses versus the corresponding calculated concentrations to obtain
a calibration relationship. Determine the best-fit straight line (y = mx + b) determined
by the method of least squares.
After the best-fit line has been drawn, determine whether the analyzer response is linear.
To be considered linear, no calibration point should differ from the best-fit line by more
than
±
2% of full scale.
If carried out carefully, the checks described in this section will provide reasonable
confidence that the T300 is operating properly. Checks should be carried out at least
every 3 months as the possibility of malfunction is always present.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...