Teledyne API – Model T300/T300M CO Analyzer
Specifications and Approvals
33
2.2.
EPA EQUIVALENCY DESIGNATION
Note
T300M: EPA equivalency does not apply to this model.
Teledyne API’s Model T300, Carbon Monoxide Analyzer, is designated as Reference
Method Number RFCA-1093-093 as defined in 40 CFR Part 53, when operated under
the following conditions:
•
Range: Any range from 10 ppm to 50 ppm
•
Ambient temperature range of 10 to 40
°
C
•
Sample filter: Equipped with 5-micron PTFE filter element in the internal filter
assembly
•
Software settings:
Dilution factor
1.0
AutoCal
ON or OFF
Dynamic Zero
ON or OFF
Dynamic Span
OFF
Dual Range
ON or OFF
Auto Range
ON or OFF
Temp/Pres Compensation
ON
Accuracy
<± (0.02% CO
2
+ 2% of reading)
Linearity
<± 0.1% CO
2
Temperature Coefficient
<± 0.01% CO
2
/°C
Rise and Fall Time
<60 seconds to 95%
1
As defined by the USEPA
2
Defined as twice the zero noise level by the USEPA
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...