Teledyne API – Model T300/T300M CO Analyzer
Getting Started
73
C
ALIBRATION
G
AS
S
OURCES
SPAN
GAS
•
Attach a gas line from the source of calibration gas (e.g. a Teledyne API’s T700E
Dynamic Dilution Calibrator) to the
SPAN
inlet.
Z
ERO
AIR
•
Zero air is supplied internally via a zero air scrubber that draws ambient air through
the IZS inlet.
I
NPUT
G
AS
V
ENTING
The zero air supply and sample gas line MUST be vented in order to ensure that the
gases input do not exceed the maximum inlet pressure of the analyzer as well as to
prevent back diffusion and pressure effects. These vents should be:
•
At least 0.2m long;
•
No more than 2m long and;
•
Vented outside the shelter or immediate area surrounding the instrument.
E
XHAUST
O
UTLET
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line
should be:
•
PTEF tubing; minimum O.D ¼”;
•
A maximum of 10 meters long;
•
Vented outside the analyzer’s enclosure.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...