Teledyne API – Model T300/T300M CO Analyzer
Calibration Procedures
221
To cause the analyzer to measure and record a value for
PRES
, press.
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
DIAG SIGNAL I/O
NEXT
ENTR
EXIT
Continue pressing NEXT until ...
EXIT discards the
new setting.
ENTR accepts the
new setting.
DIAG
PRESSURE CALIBRATION
PREV NEXT
ENTR
EXIT
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST TST> CAL
SETUP
DIAG PCAL
PRESSURE CALIBRATION
CAL
EDIT
EXIT
DIAG PCAL
SAMPLE PRESS: 29.93 IN-HG-A
2
9
.9
3
ENTR EXIT
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...