Teledyne API – Model T300/T300M CO Analyzer
A Primer on Electro-Static Discharge
331
•
Folding the open end over isolates the component(s) inside from the effects of
static fields.
•
Leaving the bag open or simply stapling it shut without folding it closed
prevents the bag from forming a complete protective envelope around the
device.
6. Once you have arrived at your destination, allow any surface charges that may have
built up on the bag or bin during travel to dissipate:
•
Connect your wrist strap to ground.
•
If you are at the instrument rack, hold the bag in one hand while your wrist strap
is connected to a ground point.
•
If you are at a anti-ESD workbench, lay the container down on the conductive
work surface
•
In either case wait several seconds
7. Open the container.
14.4.2.4.
OPENING SHIPMENTS FROM TELEDYNE API’S CUSTOMER SERVICE
Packing materials such as bubble pack and Styrofoam pellets are extremely efficient
generators of static electric charges. To prevent damage from ESD, Teledyne API ships
all electronic components and assemblies in properly sealed anti-ESD containers.
Static charges will build up on the outer surface of the anti-ESD container during
shipping as the packing materials vibrate and rub against each other. To prevent these
static charges from damaging the components or assemblies being shipped make sure
that you:
Always unpack shipments from Teledyne API Customer Service by:
1. Opening the outer shipping box away from the anti-ESD work area.
2. Carry the still sealed anti-ESD bag, tube or bin to the anti-ESD work area.
3. Follow steps 6 and 7 of Section 14.4.2.3 above when opening the anti-ESD
container at the work station.
4. Reserve the anti-ESD container or bag to use when packing electronic components
or assemblies to be returned to Teledyne API.
14.4.2.5.
PACKING COMPONENTS FOR RETURN TO TELEDYNE API’S CUSTOMER SERVICE
Always pack electronic components and assemblies to be sent to Teledyne API’s
Customer Service in anti-ESD bins, tubes or bags.
CAUTION
ESD Hazard
•
DO NOT
use pink-poly bags.
•
NEVER
allow any standard plastic packaging materials to touch the
electronic component/assembly directly.
•
This includes
,
but is not limited to, plastic bubble-pack, Styrofoam
peanuts, open cell foam, closed cell foam, and adhesive tape.
•
DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD tape.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...