SETUP
à
COMM: COMMUNICATIONS PORTS
This section introduces the communications setup menu; Section 6 provides the setup
instructions and operation information. Press SETUP>ENTR>MORE>COMM to arrive
at the communications menu.
5.7.1.
ID (MACHINE IDENTIFICATION)
Press ID to display and/or change the Machine ID, which must be changed to a unique
identifier (number) when more than one instrument of the same model is used:
•
in an RS-232 multidrop configuration (Sections 3.3.1.8 and 6.7.2)
•
on the same Ethernet LAN (Section 6.5)
•
when applying MODBUS protocol (Section 6.7.1)
•
when applying Hessen protocol (Section 6.7.2)
The default ID is the same as the model number; for the Model T100, the ID is 0100.
Press any button(s) in the MACHINE ID menu (Figure 5-2) until the Machine ID
Parameter field displays the desired identifier.
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SETUP X.
MACHINE ID: 300 ID
0 3 0 0
ENTR EXIT
Toggle
to cycle
through the available
character set:
0-9
ENTR accepts the new
settings
EXIT ignores the new
settings
Figure 5-2:
COMM– Machine ID
The ID can be any 4-digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)
5.7.2.
INET (ETHERNET)
Use SETUP>COMM>INET to configure Ethernet communications, whether manually
or via DHCP. Please see Section 6.5 for configuration details.
5.7.3.
COM1 AND COM2 (MODE, BAUD RATE AND TEST PORT)
Use the SETUP>COMM>COM1[COM2] menus to:
•
configure communication modes (Section 6.2.1
•
view/set the baud rate (Section 6.2.2)
•
test the connections of the com ports (Section 6.2.3).
Configuring COM1 or COM2 requires setting the DCE DTE switch on the rear panel.
Section 6.1 provides DCE DTE information.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...