Table A-2: T300/T300M and M300E/EM Warning Messages, Revision L.8
Name
1
Message Text
Description
Warnings
WSYSRES SYSTEM
RESET
Instrument was power-cycled or the CPU
was reset.
WDATAINIT
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT CONFIG
INITIALIZED Configuration storage was reset to factory
configuration or erased.
WCONCALARM1
CONC ALARM 1 WARN
Concentration limit 1 exceeded.
WCONCALARM2
CONC ALARM 2 WARN
Concentration limit 2 exceeded.
WSOURCE SOURCE
WARNING Reference reading minus dark offset
outside of warning limits specified by
REF_SOURCE_LIMIT
variable.
WAUTOZERO
4, 5
AZERO WARN 1.001
Auto-reference ratio below limit specified
by
ZERO_LIMIT
variable.
WBENCHTEMP
BENCH TEMP WARNING
Bench temperature outside of warning
limits specified by
BENCH_SET
variable.
WWHEELTEMP
WHEEL TEMP WARNING
Wheel temperature outside of warning
limits specified by
WHEEL_SET
variable.
WO2CELLTEMP
10
O2 CELL TEMP WARN
O
2
sensor cell temperature outside of
warning limits specified by
O2_CELL_SET
variable.
WSAMPFLOW
6
SAMPLE FLOW WARN
Sample flow outside of warning limits
specified by
SAMP_FLOW_SET
variable.
WSAMPPRESS
SAMPLE PRESS WARN
Sample pressure outside of warning limits
specified by
SAMP_PRESS_SET
variable.
WSAMPTEMP
SAMPLE TEMP WARN
Sample temperature outside of warning
limits specified by
SAMP_TEMP_SET
variable.
WPURGEPRESS
9
PURGE PRESS WARN
Purge pressure outside of warning limits
specified by
PURGE_PRESS_SET
variable.
WBOXTEMP BOX
TEMP
WARNING Internal box temperature outside of
warning limits specified by
BOX_SET
variable.
WBOXTEMP2
4
BOX
TEMP2
WARNING Internal box temperature #2 outside of
warning limits specified by
BOX2_SET
variable.
WOVENTEMP
11
OVEN TEMP WARNING
Oven temperature outside of warning
limits specified by
OVEN_SET
variable.
WPHOTOTEMP PHOTO
TEMP
WARNING Photometer temperature outside of
warning limits specified by
PHOTO_TEMP_SET
variable.
WDYNZERO
CANNOT DYN ZERO
Contact closure zero calibration failed
while
DYN_ZERO
was set to
ON
.
WDYNSPAN
CANNOT DYN SPAN
Contact closure span calibration failed
while
DYN_SPAN
was set to
ON
.
WREARBOARD
REAR BOARD NOT DET
Rear board was not detected during
power up.
06864B DCN6314
Summary of Contents for T300
Page 2: ......
Page 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Page 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Page 26: ...24 06864B DCN6314 ...
Page 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Page 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Page 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Page 244: ...06864B DCN6314 ...
Page 245: ...243 06864B DCN6314 ...
Page 246: ...06864B DCN6314 ...
Page 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Page 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Page 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Page 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Page 401: ...06864B DCN6314 D 7 ...