3700 Portable Samplers
Section 3 Programming Guidelines
3-5
Figure 3-2 Samples-Per-Bottle Multiplexing
Bottles-Per-Sample Multiplexing –
I n b o t t l e s - p e r- s a m p l e
multiplexing, you fill a programmed number of bottles at each
sample event. (The programmed number of bottles is a bottle set;
as many as 24 bottles can be used.) Bottles-per-sample multi-
plexing is useful when you need a sample that exceeds the
capacity of your bottles. It is also useful when you need sets of
identical samples; for example, when using more than one pre-
servative.
When you use bottles-per-sample multiplexing, you modify the
normal sampling cycle (relocate distributor, pre-sample purge,
sample delivery, post-sample purge) by delaying the complete
post-sample purge until sample volumes have been delivered to
each bottle of the set. Instead, the pump reverses until the
pump
tubing
is purged just beyond the liquid detector. When the
purged liquid passes the detector, the distributor moves to the
next bottle, and the pump begins pumping forward to deliver the
next sample. This repeats until each bottle of the set is filled. A
full post-sample purge is completed at the end of the sample
event. The sampler then pauses until the next sample event
occurs, at which time the cycle repeats for the following bottle
set. Figure
3-3 illustrates bottles-per-sample multiplexing.
Figure 3-3 Bottles-Per-Sample Multiplexing
3
4
5
6
7
8
1
2
Bottle Number (8 of 24 Bottles Shown)
1
2
3
4
5
6
7
8
Sample Event Number
1
1
2
2
3
3
4
4
Bottle Number (8 of 24 Bottles Shown)
1
2
3
4
5
6
7
8
Sample Event Number
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......