3700 Portable Samplers
Section 1 Introduction
1-4
In samples-per-bottle multiplexing, more than one sample
volume can be placed in a bottle. Samples-per-bottle multiplexing
allows you to collect a
series
of small composite samples.
If you deposit several samples in each bottle, the contents rep-
resent an average of the flow stream during the sampling period.
In bottles-per-sample multiplexing, you can place equal sample
volumes in as many as 24 bottles at each sample event.
Bottles-per-sample multiplexing is ideal for situations where you
need
identical sets
of samples – when you need to use more than
one preservative, for example. You can multiplex both time-paced
and flow-paced sequential samples.
The third type of multiplexing, multiple-bottle compositing, is
accessible through the extended programming mode. Mul-
tiple-bottle compositing places more than one sample volume
into more than one bottle. At each sample event, a sample
volume is placed in several bottles, in sets of as many as 24
b o t t l e s . M u l t i p l e - b o t t l e c o m p o s i t i n g c o m b i n e s
bottles-per-samples multiplexing and samples-per-bottle multi-
plexing; it is applicable to situations where you need a
series of
identical sets
of samples. You can also use multiple bottle com-
positing when you need a series of larger volume composite
samples by taking several smaller samples and distributing
them over several bottles.
In the extended programming mode, you can switch multiplexed
bottles or bottle sets after a programmed number of samples
have been deposited or after a programmed period of time. Both
methods can be used with either time-paced or flow-paced rou-
tines. By specifying the number of samples to be deposited, you
can control the volume of each bottle precisely. By specifying that
the sets be switched after specific time periods, you can control
the time frame for a series of sample volumes. This is especially
useful for flow-paced sampling. Although the flow rate may vary,
each bottle or bottle set represents a fixed time period.
1.2.3 STORM Sampling
The 3700 Series storm capabilities are ideal for monitoring storm
water runoff. The sampler allows you to divide the bottles into
two groups. The first group of bottles is normally reserved for a
first-flush grab sample. The second bottle group receives the
flow-weighted composite samples. You can distribute samples in
either group sequentially or in any of the three types of multi-
plexing. Three bottle configurations are available for STORM
sampling with the 3700 Standard are: the 24-bottle configuration
with either 1,000 or 350 ml bottles, a 12-bottle configuration con-
taining 1 quart (950 ml) glass bottles, and a 4-bottle configu-
ration with 1-gallon (3,800 ml) glass bottles. With the 3700
Compact, 24-bottle sets have a capacity of 500 ml and 12-bottle
sets have a 300 ml capacity.
STORM sampling takes full advantage of the sampler-enable
programming available through the Teledyne Isco 4200 Series
Flow Meters, or the 4100 Series Flow Logger and the Flowlink
®
software. You can program the flow meter to monitor the flow
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......