display, distributor, and pump; and lets you reinitialize RAM.
T h e d i s p l ay i n f o r m a t i o n i s d i s c u s s e d b e l o w i n o r d e r o f
appearance.
Software Revision Number –
The sampler will display the
software revision number for a short period of time.
Test RAM and ROM –
RAM and ROM test messages replaced
the software revision display. A successful test is indicated by the
messages “RAM PASSED TEST” or “ROM PASSED TEST.” If
either RAM or ROM fail the test, the sampler displays one of the
following messages: “RAM FAILED TEST” or “ROM FAILED
TEST” and beeps every three seconds until you turn it off. If
either the RAM or ROM fail the test, service the sampler.
Contact Teledyne Isco Customer Service for assistance.
LCD Test –
After successful RAM and ROM tests, the sampler
will then test the LCD by first filling the display with solid rect-
angles and then printing the alphabet and other characters.
Each character position in the display should contain a character.
Pump Test –
The next step tests the pump. During the test, the
pump will run briefly and the display will indicate an “OFF/ON”
number. The number should fall within the range of 50 to 200. If
the count falls below 50 or exceeds 200, have the pump serviced.
A count near 100 is typical.
Distributor Test –
Following the pump test, you can test the
distributor. If you elect to test the distributor, the sampler will
move the distributor to Position 24, then return it to Position 1.
The positions are indicated on the display. If the message, “DIS-
TRIBUTOR JAMMED” is displayed, check the routing of the
pump tubing and the bottles in the tub. If the tubing and bottles
are correctly positioned, the distributor drive mechanism may
need servicing. (
Display #370
)
Reinitialize Controller –
In the final steps you reinitialize the
sampler. If you select “NO,” the sampler will return to the con-
figure option list. If you select “YES,” the sampler will reset a
number of configuration and program settings, then you can turn
off the sampler. (
Display #371
)
Table
3-4 lists the reinitialized settings.
Not all settings are
reset.
Set Clock, Bottle Size, Suction Line, and Sampler ID con-
figure-option settings remain unchanged. This reduces the
number of settings you would need to change if the sampler were
accidently reinitialized. The Pump Count total is not reset to
maintain an accurate count for the Tubing Life Warning.
3.9.21 Exit Configuration
This option allows you to leave the configure sequence and
return to the standby state. There are no input displays. Press
the
ENTER
/
PROGRAM
key to exit the configure sequence. You can
exit the configuration sequence at any time using the
EXIT
PROGRAM
key.
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......