4-1
3700 Portable Samplers
Section 4 Options and Interfacing Equipment
4.1 Introduction
This section discusses major options available with the Teledyne
Isco 3700 Standard and Compact Samplers. Included is infor-
mation about:
• Conversion for composite sampling
• Programming and configuring the controller for
composite sampling
• Connection to Teledyne Isco and non-Isco Flow meters
• Interface devices
• Master/Slave operation
• Use with the 1640 Liquid Level Actuator
4.2 Composite Sampling
The 3700 Standard and Compact Samplers are fully program-
mable for time-paced and flow-paced composite sampling. The
object of composite sampling is to collect a series of discrete,
equal-volume samples at known time or flow intervals, and to
place these samples into a single composite sample container.
4.2.1 Conversion Procedure
To prepare your 3700 Standard for composite sampling, you must
convert the center section. To do so, remove the distributor arm
assembly, install a shorter pump tube, and install the optional
composite shut-off float cage assembly. With the portable, com-
plete the preparation by the adding a composite sample con-
tainer and locating ring in the standard bottle base or an
optional composite base. With the compact, remove the inner
sleeve and install the proper bottle configuration for composite
sampling.
These are the listed steps needed to convert the 3700:
1. Separate the center section from the remainder of the sam-
pler.
2. Remove the pump tube and distributor arm from the sam-
pler, as described in
Removing the Pump Tube
(Section 5).
3. Turn the center section over, as shown in Figure 4-1, and
install the optional float and float cage. You should install
the shaft on the “diving-bell” float in the hole in the distrib-
utor shaft, as shown. Attach the float cage to the underside
of the center section using the four screws and the 1-inch
spacers provided with the float-cage kit.
When attaching the cage, reuse the existing lockwash-
ers.Be sure the float cage is oriented with the tube guide
positioned as shown in Figure 4-1. The float and float cage
direct the discharge of the sample into the composite con-
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......