3700 Portable Samplers
Section 3 Programming Guidelines
3-82
Example
Time-Paced Multiplexed (samples-per-bottle) Sampling
shows how to
program the sampler to take time-paced samples at
uniform time intervals of 15 minutes. At each sample event, one
bottle will receive a sample volume of 75 ml. The bottles are to be
switched every hour so that four sample volumes have been
deposited into each bottle. The sampling routine will start
according to the programmed start-time delay set in the Start
Time Delay configure option.
BOTTLE 1
PUMPING 500 ml
This display appears when the post-sample purge from the previous sample is completed. It indicates the bottle
number which is to receive the sample at the upcoming sample event. The second line displays the scheduled time for
the next event on the left. The current time is displayed on the right.
BOTTLE 2
AT 8:30 8:01:11
The cycle of displays is repeated for each bottle until the sampling routine is done.
↓
↓
When the routine is completed, this message appears. It reports the status of the routine (‘DONE”), the total number
of sample events, and the current time and date.
DONE. . . 24 SAMPLES
8:10:35 21-JUN-04
Example: Time-Paced Sampling: Uniform Intervals (Continued)
Example: Time-Paced Multiplexed (samples-per-bottle) Sampling
1.
Press
ENTER
/
PROGRAM
to access the interactive state.
. . . STANDBY . . .
5:37:23 19-JUN-04
2.
Access the program sequence by selecting “PROGRAM.”
[
PROGRAM
, CONFIGURE]
SAMPLER
3.
Select “TIME.”
[
TIME
, FLOW, STORM]
PACED SAMPLING
4.
Select “UNIFORM.”
[
UNIFORM
, NONUNIFORM]
TIME INTERVALS
5.
Enter “0” to set the hours at zero. Press
ENTER
/
PROGRAM
to store the number “0” and move to the minutes entry.
SAMPLE EVERY
0 HOURS,30
MINUTES
6.
Enter “15” to set the minutes entry to 15.
SAMPLE EVERY
0 HOURS,
15
MINUTES
7.
Enter “1” to set the number of bottles per sample event to 1.
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......