3.5.3 Configure Sequence
The program sequence and the configure sequence are interde-
pendent. The selections you make in the configure sequence
determine what settings will be available to you in the program
sequence. When you select the extended programming mode, the
program sequence is modified to accommodate the extended fea-
tures by adding one or more input displays.
The extended programming mode not only extends the number of
features available in the program sequence, it extends the
number of configure options. When you select extended pro-
gramming, for example, you can enable the Sampling Stop and
Resumes and/or Nonuniform Times features so you can use them
in the extended program sequence.
When you select the basic programming mode, these configure
options are not available to you. As in the programming
sequence, less frequently used options have been separated to
simplify the configuration process.
Most configure options are available in both programming
modes. Table
3-1 summarizes the configure sequence options;
each option is marked as available in both modes or available
only in the extended mode. A detailed discussion of each configu-
ration option is found in the sections beginning with
Set Clock
through
Exit Configuration
.
It is usually not necessary to change the configure option settings
unless the sampling application changes significantly. The
sampler is
always
programmed and configured: the sampler’s
internal battery maintains program and configure settings. In
fact, the sampler is shipped with factory program and configure
settings. It is configured for the basic program mode and for the
bottle configuration and suction line ordered with the sampler.
3.6 Programming
Procedure
The procedure used to program the sampler follows the steps
listed below.
Note
Because the configuration settings determine portions of the
program sequence and affect the accuracy of the sample,
check the configuration settings before programming.
1. Determine the equipment you will be using with the sam-
pler. You will need to know the capacity of the bottles and
the inside diameter, type (vinyl or TEFLON), and length of
the suction line. You will need this information to verify
and revise the configuration settings in step 2.
If you will be using the sampler for composite sampling,
you may need to convert the sampler. The section
Conver-
sion for Composite Sampling
in Section
4.2 provides infor-
mation on converting the 3700 Standard for composite
sampling. Example
onstrates the configuration changes you must make.
2. Check the configuration settings. From standby, access the
interactive state. Select the configure sequence. Check the
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......