3700 Portable Samplers
Section 4 Options and Interfacing Equipment
4-5
Teledyne Isco also offers a connector prewired to a 22-foot
(6.7-meter) cable that terminates in two wires. The black wire is
connected to pin A; the white wire to pin C.
Note that you will be unable to communicate anything other
than the flow pulse contact closure with a non-Isco Flow meter.
The non-Isco Flow meter will not be able to interpret event and
bottle number information.
If the flow pulse generated by the contact closure on the flow
meter is not compatible with Teledyne Isco’s standard, contact
Teledyne Isco’s Special Products Department for an appropriate
interface device.
4.3.3 4-20 mA Flow Signals
The 3700 Sampler can also be used with flow meters having
other types of outputs, for example, a 4 to 20 mA output directly
proportional to the flow rate. However, these flow meters require
a special interface device to convert their output signal to one
compatible with the 3700 sampler.
The
4–20 mA Sampler Input Interface
converts 4 to 20 mA
output signals from both non-Isco Flow meters and closed-pipe
flow meters to flow pulses acceptable to the 3700 Sampler.
Figure 4-3 4-20 mA Sampler Input Interface
4.3.4 1640 Liquid Level
Actuator
The Liquid Level Actuator is an accessory that begins a sampling
routine when the liquid level of the flow stream reaches a prede-
termined height. The actuator inhibits sampling according to
flow levels; that is if the level of the flow stream falls below the
actuator’s probe, the actuator will inhibit sampling until the level
again rises to contact the probe.
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......