3700 Portable Samplers
Section 4 Options and Interfacing Equipment
4-3
Prepare the tub for composite sampling by installing the sample
container. No locating deck is required for the composite tub:
locating depressions are integrally molded into the tub. You can
then attach the composite base to the center section. Note: When
cooling the samples with ice, the melt water may cause the bottle
to float when you remove the center section.
4.2.4 Programming and
Configuring for
Composite Sampling
After converting the sampler, be sure to change the bottle
number and size settings in the Bottles and Sizes configure
option. (Refer to
Event Mark and Bottle Signal Output
.)
You
must reprogram the controller for composite sampling because
changing the bottle setting will change the program sequence.
Refer to
Programming Introduction
for information on pro-
gramming the controller.
4.3 Connections to
External Devices
The 3700 Series Samplers can collect samples on a flow-paced
basis using flow inputs from an external flow meter. The 3700
requires a 5-to-15 VDC pulse of at least 25-millisecond duration
to register a flow pulse.
Open-channel
flow meters are con-
nected to the sampler by attaching a flow meter connect cable to
the flow meter connector located on the rear of the sampler.
Closed-pipe
flow meters are connected to the sampler’s Flow
Meter port using the Teledyne Isco 4-20 mA Input Interface
device (see Appendix B Accessories).
Your sampler has an additional connector, labeled “PRINTER,”
used to connect the sampler to an Isco Field Printer or laptop
computer. Both the flow meter connector and the printer con-
nector use the same 6-pin configuration. The printer cable con-
nector accepts both the connector from the field printer and a
cable from a field computer or a laptop computer running the
Samplink software.
Note
The Field Printer has been discontinued and is no longer sold
by Teledyne Isco.
4.3.1 Isco Flow Meters and
Flow Loggers
The 4200 Series Flow Meters, 2100 Series Flow Modules, and
4100 Series Flow Loggers, and UniMag Closed-Pipe Flow Meters,
are collectively called “flow meters” in this manual. You can pur-
chase cables to connect the 3700 samplers to an Isco flow meter,
flow logger, or 2100 series flow module. Refer to the
Accessories
List
for details. To make the connection, attach the appropriate
cable connector to the flow meter according to directions in the
flow meter in instruction manual, attach the other connector to
the 6-pin flow meter connector on the rear of the sampler.
Note
If a connect cable is not attached to the flow meter or print con-
nectors, be sure that the protective covers are tightly fastened in
place.
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......