3700 Portable Samplers
Section 3 Programming Guidelines
3-34
Example
demonstrates the pro-
gramming procedures for a composite sampling routine. When
composite sampling with the 3700 Standard, you must convert
the unit for single-bottle sampling. With the 3700 Compact, you
must install the proper bottle into the base.
After converting the 3700 Standard, check the settings to verify
the sampler is configured for composite sampling. This example
begins with a demonstration of the steps needed to check the
Bottles and Sizes configure option settings. The Bottles and Sizes
configure settings should indicate that only one bottle is being
used. Be sure to enter the correct bottle volume. The settings
used in this example are for the 4-gallon (15,000 ml) composite
bottle available with the 3700 Standard.
This time-paced routine will take samples at 15-minute intervals
for a 12-hour period. Sample volumes are to be 250 ml each.
When entering the program settings, you must enter the number
of samples required before entering the sample volumes. At four
samples per hour, the 12-hour period would yield 48 samples.
The total sample volume collected at the end of the routine would
be 12,000 ml (48 samples
x
250 ml/sample), well within the
capacity of the 4-gallon bottle.
Example: Time-Paced Composite Sampling
1.
If the sampler is not already on, press the
ON
/
OFF
key to turn it on. Press
ENTER
/
PROGRAM
to access the
interactive state.
. . . STANDBY . . .
10:34:50 19-JUN-04
2.
Access the configure sequence by selecting “CONFIGURE.”
[PROGRAM,
CONFIGURE
]
SAMPLER
3.
Press the
LEFT
ARROW
or
RIGHT
ARROW
key to scroll through the configure options until the Bottles and Sizes
configure option appears.
SELECT OPTION: (
← →
)
SET CLOCK
4.
Press the
ENTER
/
PROGRAM
key to access the Bottles and Sizes settings.
SELECT OPTION: (
← →
)
BOTTLES AND SIZES
5.
Select “PORTABLE” for either the 3700 Standard or Compact.
[
PORTABLE
, REFRIG
SAMPLER
6.
Select “1” for single bottle composite sampling.
[
1
, 4, 12, 24]
BOTTLES
7.
Enter the bottle size here, “15,000.” (
Table
3-2 lists the standard Teledyne Isco bottle sizes.)
BOTTLE VOLUME IS
15000
ml
8.
Press the
ENTER
/
PROGRAM
key at this display to access the Suction Line input displays.
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......