3700 Portable Samplers
Appendix D Display Index
D-9
Enable Pin configure option. Select “YES” to turn the Master/slave mode on. Select “NO” to turn the Master/Slave
mode off.
Display 320
MASTER/SLAVE MODE?
[YES, NO]
Enable Pin configure option. Select “YES” to take a sample at the time the sampler becomes disabled. Select “NO” if
you do not want a sample event when the sampler becomes disabled.
Display 321
SAMPLE UPON DISABLE?
[YES, NO]
Enable Pin configure option. Select “YES” to take a sample when the sampler becomes enabled;
i.e.,
ceases to
receive an inhibit signal from a flow meter or a Liquid Level Actuator. Note: No sample will be taken if the enable sig-
nal is used to initiate the start-time delay countdown. Select “NO” if you do not want a sample event when the sampler
becomes enabled.
Display 322
SAMPLE UPON ENABLE?
[YES, NO]
Enable Pin configure option. Select “YES” to restart the sample interval countdown at the time the sampler is
enabled. The interval countdown is determined by the setting if “NO” is selected in Display #324.
Display 323
RESET SAMPLE
INTERVAL? [YES, NO]
Enable Pin configure option. This display appears if you select “NO” in Display #323. Select “YES” to freeze the sam-
ple interval when the sampler becomes disabled. When the sampler is enabled, the interval countdown continues
from the point where it had stopped. Select “NO” to let the sample interval countdown continue while the sampler is
disabled.
Display 324
INHIBIT COUNTDOWN?
[YES, NO]
Event Mark configure option. Select “CONTINUOUS SIGNAL” to send a variable-duration event mark signal from pin
E of the flow meter connector. Select “PULSE” to send a 3-second event mark signal.
Display 330
[CONTINUOUS SIGNAL,
PULSE]
Event Mark configure option. This setting appears when you select “CONTINUOUS SIGNAL” in Display #330. Select
“PUMP CYCLE” to transmit the event mark signal during the entire pump cycle, including any programmed rinses and
retries. Select “FWD PUMPING ONLY” to send the signal when the pump is delivering a sample volume only.
Display 331
DURING [PUMP CYCLE,
FWD PUMPING ONLY]
Event Mark configure option. This setting appears when you select “PULSE” in Display #330. Select “PURGE” to
transmit an event mark signal to a flow meter at the beginning of the pre-sample purge. Select “FWD PUMPING” to
transmit a signal when the pump switches forward to deliver the sample volume.
Display 332
AT THE BEGINNING OF
[PURGE, FWD PUMPING]
Purge Counts configure option. Enter the number of pre-sample pump counts needed to purge the suction line. This
value is set to 150 when the controller is reinitialized.
Display 340
--- PRE-SAMPLE
COUNTS (0 - 9999)
Purge Counts configure option. Enter the number of post-sample pump counts needed to purge the suction line. The
Controller derives the number that initially appears in this display from the suction line ID and length entered in Dis-
play #230 and #232.
Display 341
--- POST-SAMPLE
COUNTS (0 - 9999)
Table D-1 Display Index (Continued)
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......