This display is part of the calibration sequence. Press the
MANUAL
SAMPLE
key when a collection bottle is underneath
the distributor and the suction line is in place.
Display 81
PRESS MANUAL SAMPLE
KEY WHEN READY...
This display is part of the calibration sequence. When it first appears, it reports the programmed sample volume. If
the measured volume differs from the reported volume, enter the measured volume here.
Display 82
--- ml VOLUME
DELIVERED
This display is part of the calibration sequence and appears if the measured volume and the programmed volume dif-
fer by a factor of two or more. Select “YES” to confirm the number entered in Display #82. Select “NO” to return to
Display #82.
Display 83
---ml! ARE YOU
SURE? [YES, NO]
Select “YES” to enter a start time. “NO” will begin the sampling routine according to the delay set in Display #310.
90
ENTER START TIME?
[YES, NO]
This display appears when you have selected “YES” in Display #90. Enter the start time and date for the first sample
event. This display will also appear if you have initiated the routine after a programmed start time so that you can
reprogram the start time.
Display 91
TAKE FIRST SAMPLE AT
HH:MM DD-MMM
This display appears when you have selected “YES” in Display #90. Enter the start time and date for the flow -pulse
countdown. This display will also appear if you have initiated the routine after a programmed start time so that you
can reprogram the start time.
Display 92
START FLOW COUNT AT
HH:MM DD-MM
This display is the last display of the storm branch. It appears when you select “YES” in Display #90. Enter the start
time for the storm-sampling routine. The sampler will disregard enable or disable signals from the flow meter until the
start time you enter here.
Display 93
STORM ENABLED AFTER
HH:MM
DD-MM
This display appears when you have selected “TIME” in either Display #40 or 41. Select “YES” if you want to specify
a clock time for the first bottle or bottle set switch. Select “NO” if you want to begin the switch interval at the routine's
start time.
Display 95
ENTER FIRST SWITCH
TIME? [YES, NO]
This display appears when you select “ENABLE” in Display #300. Enter the number of stop and resume times.
Display 100
-- STOP or RESUME
TIMES (0 - 24)
This display appears when the setting in Display #100 is greater than zero. Enter the appropriate stop time.
Display 101
STOP SAMPLING AT
1.
HH:MMDD-MMM
This display appears when the setting in Display #100 is greater than one. It follows Display #101. Enter the appropri-
ate resume time.
Display 102
RESUME SAMPLING AT
1.
HH:MMDD-MMM
Table D-1 Display Index (Continued)
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......