3700 Portable Samplers
Section 3 Programming Guidelines
3-90
TAKE
24
SAMPLES
(1-999)
7.
Enter the number of samples to occur at the first interval. The sampler is to take samples at 10-minute intervals
for an hour or 6 samples in the first hour. Since the first sample is taken at the start time, the remaining five
samples will occur at the 10-minute interval.
QUANTITY AT INTERVAL
1.
5 AT 10 MINUTES
8.
When you have completed the first entries, the display will “roll” to move the first interval entries to the first line,
and add the second line entries for the next series of samples and intervals. Enter the next entries: “3” samples
at“20” minutes.
1.
5 AT 10 MINUTES
2.
3 AT 20 MINUTES
9.
Enter the third set of samples and intervals, “15” samples at “30” minutes.
2.
3 AT 20 MINUTES
3.
15 AT 30 MINUTES
10. Enter “1” to set the number of bottles per sample event to 1.
1
BOTTLES PER
SAMPLE EVENT (1 - 24)
11. Select “SAMPLES.”
CHANGE BOTTLES BASED
ON [TIME,
SAMPLES
]
12. Enter “1” to set the number of samples per bottle to 1.
CHANGE BOTTLES EVERY
1
SAMPLES (1 - 50)
13. Enter “500” to set the sample volume at 500 ml.
SAMPLE VOLUMES OF
500
ml (10 - 1000)
14. Select “NO.”
ENTER START TIME?
[YES,
NO
]
After this message is displayed briefly, the sampler will automatically return to the standby state.
PROGRAMMING SEQUENCE
COMPLETE
15. Press the
START
SAMPLING
key to run the program.
. . . STANDBY . . .
10:37:23 19-JUN-04
16. To start the sampling routine with the first bottle, accept the blinking “1” by pressing the
ENTER
/
PROGRAM
key. If
you want to start the routine with another bottle, enter the starting bottle number here.
START SAMPLING
BOTTLE
1
(1-24)
This display appears while the Actuator is inhibiting the sampler. The second line indicates the current time and date.
SAMPLER INHIBITED!
10:35
20-JUN-04
This display appears when the sampler becomes enabled and counts down the time remaining to the start time. The
first line reports the bottle which will receive the next sample volume. The second line reports the start time at the left
and the current time on the right.
Example: Time-Paced Samples at Nonuniform Intervals (Continued)
Summary of Contents for 3700
Page 2: ......
Page 4: ......
Page 33: ...3700 Portable Samplers Section 2 Setup Procedures 2 9 Figure 2 7 Locked Sampler ...
Page 34: ...3700 Portable Samplers Section 2 Setup Procedures 2 10 ...
Page 144: ...3700 Portable Samplers Section 3 Programming Guidelines 3 110 ...
Page 152: ...3700 Portable Samplers Section 4 Options and Interfacing Equipment 4 8 ...
Page 164: ...3700 Portable Samplers Section 5 Routine Maintenance 5 12 ...
Page 180: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 2 ...
Page 181: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 3 ...
Page 182: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 4 ...
Page 183: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 5 ...
Page 184: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 6 60 3703 262 REV J 3 2 ...
Page 186: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 8 ...
Page 188: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 10 ...
Page 189: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 11 ...
Page 190: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 12 ...
Page 191: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 13 ...
Page 192: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 14 ...
Page 193: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 15 ...
Page 194: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 16 ...
Page 195: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 17 ...
Page 196: ...3700 Portable Samplers Appendix A Replacement Parts Lists A 18 ...
Page 200: ...3700 Portable Samplers Appendix B Accessories B 4 ...
Page 210: ...3700 Portable Samplers Appendix C General Safety Procedures C 10 ...
Page 224: ...3700 Portable Samplers Appendix E Calculating Flow Increment Between Samples E 4 ...
Page 234: ...3700 Portable Samplers Appendix G Material Safety Data Sheets G 8 ...
Page 238: ......
Page 240: ......
Page 242: ......